ICRA 2015poster30 citations

Toward tissue penetration by MRI-powered millirobots using a self-assembled Gauss gun

Aaron T. Becker, Ouajdi Felfoul, Pierre E. Dupont

Abstract

MRI-based navigation and propulsion of millirobots is a new and promising approach for minimally invasive therapies. The strong central field inside the scanner, however, precludes torque-based control. Consequently, prior propulsion techniques have been limited to gradient-based pulling through fluid-filled body lumens. This paper introduces a technique for generating large impulsive forces that can be used to penetrate tissue. The approach is based on navigating multiple robots to a desired location and using self-assembly to trigger the conversion of magnetic potential energy into sufficient kinetic energy to achieve penetration. The approach is illustrated through analytical modeling and experiments in a clinical MRI scanner.

BibTeX
@inproceedings{icra2015_towardtissuepene,
  title = {Toward tissue penetration by MRI-powered millirobots using a self-assembled Gauss gun},
  author = {Aaron T. Becker and Ouajdi Felfoul and Pierre E. Dupont},
  booktitle = {ICRA 2015},
  year = {2015}
}
Toward tissue penetration by MRI-powered millirobots using a self-assembled Gauss gun · ICRA 2015