ICRA 2026poster0 citations
Local Linearized Cosserat Rod Model for Contact Force Estimation in Flexible Medical Instruments and Continuum Robots
Christoph Eyberg, Johannes Horsch, Thomas Bauernhansl, Jens Langejürgen
Abstract
Knowledge of instrument contact forces can lead to safer medical interventions. We present a formulation of the frequently used Cosserat rod model that is linearized around the measured shape for efficient model-based contact force estimation in flexible instruments and robots. Validation on instruments’ deflection in an endovascular use case resulted in an average force estimation error of only 14 %.
Modeling, Control, and Learning for Soft RobotsMedical Robots and SystemsForce and Tactile Sensing