Breaking the Latency Barrier: Synergistic Perception and Control for High-Frequency 3D Ultrasound Servoing
Yizhao Qian, Yujie Zhu, Jiayuan Luo, Li Liu, Yixuan Yuan, Hongen Liao, Guochen Ning
Abstract
Tracking moving anatomical targets with robotic ultrasound is particularly challenging when the target motion is both fast and large in scale, as the end-to-end latency of existing systems prevents the perception–control loop from closing fast enough. In this paper, we argue that overcoming this limitation calls for the joint design of perception and control, rather than optimizing each in isolation. We present a tightly-coupled framework with two main components: (1) a Decoupled DualStream Perception Network that estimates 3D translational state from 2D ultrasound images at high frequency, and (2) a Single-Step Flow Policy that outputs an entire action sequence in one forward pass, removing the need for iterative rollouts used in conventional policies. Together, the two modules enable closed-loop control at over 60 Hz. In phantom experiments with complex 3D trajectories, the system achieves a mean tracking error below 6.5 mm and re-acquires the target after resultant displacements exceeding 170 mm. It tracks targets moving at speeds up to 102 mm/s with a terminal error under 1.7 mm. In-vivo trials on a human volunteer further confirm that the approach transfers to realistic clinical conditions. To our knowledge, this is the first RUSS framework to unify high-bandwidth dynamic tracking with large-scale repositioning within a single architecture, offering a concrete step toward autonomous ultrasound operation in the presence of patient motion.